Mini

The Mini TPA/PD is a compact autocorrelator designed for routine tasks. Its compact dimensions make it easy to carry and suitable for smaller spaces. The Mini provides the perfect combination of autocorrelation measurement without tuning and high sensitivity.

Key features

  • Wide range of optical kits for spectral coverage from 340 nm to 3200 nm
  • UV measurements without autocorrelation with Mini TPA
  • Short pulse measurements up to 10 fs with Mini PD
  • Compact design
  • Ultra-precise delay resolution
  • Software and USB interface
  • TCP/IP remote control with standardized command set for easy programming
  • Calibration with NIST compliance
  • Aluminum carrying case for easy transport
Martin Klečka
Expert advisor

Ing. Martin Klečka

+420 607 014 278

klecka@optixs.cz

Send inquiry

OptiXs care

  • We will expertly consult your planned application
  • Our team is able to integrate the product into a larger system
  • We ensure fast delivery of spare parts and local service
What else we can help with

Popis produktu

Wavelength adjustment with Mini TPA without tuning

Autocorrelators typically use a method of splitting the laser beam into two beams and then recombining them to generate the second harmonic (SHG) frequency in a nonlinear crystal. APE's Mini TPA autocorrelator works on a different principle - absorption of two photons. This eliminates the need for angular tuning of the SHG crystal and makes the wavelength tuning process unnecessary.

Switching between collinear and non-collinear modes using Mini PD

Mini PD supports fast and easy switching between collinear and non-collinear measurement modes. While collinear mode provides pulse length measurements with additional qualitative information about the pumping beam, non-collinear mode provides background-free autocorrelation with high dynamic range.

UV range without cross-correlation

Together with the UV Optical Kit, the Mini TPA provides simple pulse width measurements in the UV range from 250 nm to 400 nm - without the need for cross-correlation. Eliminating cross-correlation also shortens the data evaluation process and increases its efficiency.

UV @ 250 nm to IR @ 3200 nm

APE provides a range of interchangeable optical sets from UV @ 250 nm to IR @ 3200 nm for sensitive measurements over a wide wavelength range. With its compact size, the Mini TPA is also the perfect answer to your requirements for space saving and easy portability

Parameters

Model TPA PD
Pulse Width 35 fs...3.5 ps 10 fs...3.5 ps
Wavelength range 250 nm...3200 nm
Depending on the optical set
420 nm...1800 nm
Depending on the optical set
Recommended repetition rate

>300 Hz

For UV only:

>300 Hz
Sensitivity* (typical)

0.1 W2
Depending on the optics set
UV only:

Depending on optics set 1 W2
Max. power input, pulsed energy 0.5 W for quazi-cw lasers

5µ for kHz laser

Input beam polarization Linear, any Linear, horizontal
Resolution delay 1 femtosecond

* Measured sensitivity including optics set, defined as average power multiplied by peak power of incident pulses: PAV * Ppeak.