We distribute this product only in the Czech Republic and Slovakia.

SuSi flowmodules

The Spetec laminar flow module(flowmodules) allows you to equip your workplace withcleanroom conditions in a simple and cost-effective way. Simply place the module in the space directly above the workstation or above the machine in operation. Ambient air is drawn in by means of a radial fan, i.e. downward flowing air passes through the filter in parallel flow lines.

The SuSi (Super Silent) variant is characterised by very quiet operation and an electronic control interface.

Key features

  • Very quiet operation, max. 52 dB
  • Night maintenance mode
  • 8 adjustable power levels
  • Ensures environmental cleanliness up to DIN ISO 5
  • Inlet filter efficiency H 14 up to 99.995%
  • Can be directly hung from the ceiling or mounted on a support structure (mounting rails included)
Petr Březina
Expert advisor

Ing. Petr Březina

+420 607 014 272

brezina@optixs.cz

Send inquiry

OptiXs care

  • We will expertly consult your planned application
  • Our team is able to integrate the product into a larger system
  • We ensure fast delivery of spare parts and local service
What else we can help with

Filtration

The basic principle of the module is simple. The air around the module is sucked in by a radial fan through the cleaning filters and then driven down in parallel jets. Dust particles that are not captured by the inlet filters are entrained by the air stream and driven away from the space.

In addition, due to the resulting overpressure inside the box or otherwise enclosed space, it is not possible for additional dust particles to enter the space.

  • Two to four pre-filters depending on the size of the module
  • Particulate filter class H14: According to the MPPS (Most Penetrating Particle Size) method, 99.995% of particles > 0.12 μm are impermeable and 99.9995% of particles > 0.3 μm are impermeable
  • Easily replace the filter from the bottom (inside the cleanroom) via a cassette system containing the main filter. If necessary, the conventional method of changing the filter from above (outside the clean room) can still be used.
  • The insulation factor is 104 - Under the laminar flow module, the air quality is 10000 times better than in the immediate surroundings. For example: an immediate environment of 1000000 particles / m³ is equal to
  • Cleanroom class 100 according to US Fed. 209E, or ISO 5 according to DIN ISO 14644-1 and better (in the space below the flow module)

Control panel

  • Optional external control panel on a 5 m cable
  • 8 airflow setting stages, reproducible values by flow rate indication on display
  • including eco-mode (night), cleaning function and operating hours counter
  • Fault indication and filter charging
  • Light on/off control (optional)
  • Possibility to control up to 100 flowmodules at a time

Parameters

SuSi model Main filter dimensions Number of pre filters Weight
FMS 37 610 x 610 mm 2 28 kg
FMS 56 915 x 610 mm 2 31 kg
FMS 75 1220 x 610 mm 4 49 kg
FMS 93 1525 x 610 mm 4 56 kg
FMS 112 1830 x 610 mm 4 63 kg